Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-61
:MEASure:ITEM
Syntax :MEASure:ITEM <item>[,<src>[,<src>]]
:MEASure:ITEM? <item>[,<src>[,<src>]]
Description Measure any waveform parameter of the specified source, or query the measurement
result of any waveform parameter of the specified source.
Parameter
<item> Discrete {VMAX|VMIN|VPP|VTOP|VBASe|VAMP|VAVG|
VRMS|OVERshoot|PREShoot|MARea|MPARea|
PERiod|FREQuency|RTIMe|FTIMe|PWIDth|
NWIDth|PDUTy|NDUTy|RDELay|FDELay|
——
<src> Discrete {CHANnel1|CHANnel2|CHANnel3|CHANnel4}
——
Explanation The [,<src>[,<src>]] parameter sets the source of the parameter to be measured.
If the parameter to be measured needs single source (such as VMAX, VMIN, VPP,
VTOP, VBASe, VAMP, VAVG, VRMS, OVERshoot, MARea, MPARea, PREShoot, PERiod,
FREQuency, RTIMe, FTIMe, PWIDth, NWIDth, PDUTy, NDUTy, etc.), one source
requires to be set. If the [,<src>[,<src>]] parameter is omitted, the default source is
selected by the
:MEASure:SOURce command.
If the parameter to be measured needs two sources (such as RDELay, FDELay,
RPHase, FPHase, etc.), two sources need to be sent with the command. If else, the
command will be invalide. If the [,<src>[,<src>]] parameter is omitted, the default
sources are selected by
:MEASure:SETup:DSA and :MEASure:SETup:DSB
or :MEASure:SETup:PSA and :MEASure:SETup:PSB.
Return
The query returns the measurement result in scientific notation.
Example
:MEASure:ITEM OVERshoot,CHANnel2 /*Enable the overshoot measurement of CH2*/
:MEASure:ITEM? OVERshoot,CHANnel2 /*The query returns 8.888889e-03*/
Related
Command
:MEASure:SOURce
:MEASure:SETup:DSA
:MEASure:SETup:DSB
:MEASure:SETup:PSA