Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-27
:CURSor:XY:AYValue?
Syntax
:CURSor:XY:AYValue?
Description Query the Y value of cursor A in XY cursor measurement mode. The unit dependents on
the amplitude unit selected by the corresponding channel.
Return
Format
The query returns the Y value of cursor A in scientific notation.
Example
:CURSor:XY:AYValue? /*The query returns 4.000000e-01*/
Related
Command
:CHANnel<n>:UNITs
:CURSor:XY:BXValue?
Syntax
:CURSor:XY:BXValue?
Description Query the X value of cursor B in XY cursor measurement mode. The unit dependents on
the amplitude unit selected by the corresponding channel.
Return
Format
The query returns the X value of cursor B in scientific notation.
Example
:CURSor:XY:BXValue? /*The query returns -4.200000e-01*/
Related
Command
:CHANnel<n>:UNITs
:CURSor:XY:BYValue?
Syntax
:CURSor:XY:BYValue?
Description Query the Y value of cursor B in XY cursor measurement mode. The unit dependents on
the amplitude unit selected by the corresponding channel.
Return
Format
The query returns the Y value of cursor B in scientific notation.
Example
:CURSor:XY:BYValue? /*The query returns -4.000000e-01*/
Related
Command
:CHANnel<n>:UNITs