Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-91
:TRIGger:SLOPe:WHEN
Syntax :TRIGger:SLOPe:WHEN <when>
Set or query the trigger condition in slope trigger.
Parameter
{PGReater|PLESs|NGReater|NLESs|PGLess|NGLess}
Explanation PGReater/PLESs: you need to specify a time value (refer to :TRIGger:SLOPe:TIME).
The oscilloscope triggers when the positive slope time of the input signal is
greater/lower than the specified time.
NGReater/NLESs: you need to specify a time value (refer to
:TRIGger:SLOPe:TIME).
The oscilloscope triggers when the negative slope time of the input signal is
greater/lower than the specified time.
PGLess/NGLess: you need to specify a upper limit (refer to
:TRIGger:SLOPe:TUPPer)
and a lower limit (refer to
:TRIGger:SLOPe:TLOWer) of time. The oscilloscope
triggers when the positive/negative slope time of the input signal is greater than the
specified lower limit and lower than the specified upper limit.
Return
The query returns PGR, PLES, NGR, NLES, PGL or NGL.
:TRIGger:SLOPe:WHEN NLESs /*Set the trigger condition to NLESs*/
Related
Command
:TRIGger:SLOPe:TIME
:TRIGger:SLOPe:TUPPer
:TRIGger:SLOPe:TLOWer
:TRIGger:SLOPe:TIME
Syntax :TRIGger:SLOPe:TIME <time>
Set or query the time value in slope trigger. The default unit is s.
Parameter
<time> Real 8ns to 10s
PGReater, NGReater: 1μs
Explanation This command is available when the trigger condition (refer to :TRIGger:SLOPe:WHEN) is
PGReater, PLESs, NGReater and NLESs.
Return
The query returns the time value in scientific notation.
Example
:TRIGger:SLOPe:TIME 0.000003 /*Set the time value to 3μs*/
:TRIGger:SLOPe:TIME? /*The query returns 3.000000e-06*/
Related
:TRIGger:SLOPe:WHEN