Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-59
:MEASure:STATistic:DISPlay
Syntax :MEASure:STATistic:DISPlay <bool>
:MEASure:STATistic:DISPlay?
Description
Enable or disable the statistic function, or query the status of the statistic function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation When the statistic function is enabled, the oscilloscope makes statistic and displays the
statistic results of at most 5 measurement items that are turned on last.
Return
Format
The query returns 0 or 1.
Example
:MEASure:STATistic:DISPlay ON
Related
Command
:MEASure:STATistic:MODE
:MEASure:STATistic:RESet
:MEASure:STATistic:ITEM
:MEASure:ITEM
:MEASure:STATistic:MODE
Syntax :MEASure:STATistic:MODE <mode>
:MEASure:STATistic:MODE?
Description
Set or query the statistic mode.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{DIFFerence|EXTRemum}
EXTRemum
Explanation DIFFerence: select difference mode. The statistic results contain current value,
average value, standard deviation and counts.
EXTRemum: select extremum mode. The statistic results contain current value,
average value, minimum and maximum.
Sending the
:MEASure:STATistic:DISPlay command can enable the statistic function.
When the statistic function is enabled, the oscilloscope makes statistic and displays
the statistic results of at most 5 measurement items that are turned on last.
Return
Format
The query returns DIFF or EXTR.
Example
:MEASure:STATistic:MODE DIFFerence /*Select difference mode*/
:MEASure:STATistic:MODE? /*The query returns DIFF*/
Related
Command
:MEASure:STATistic:DISPlay
:MEASure:STATistic:RESet
:MEASure:STATistic:ITEM