Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-55
:MEASure:ADISplay
Syntax :MEASure:ADISplay <bool>
Description Enable or disable all measurement function, or query the status of all measurement
Parameter
Explanation The all measurement function can measures the following 20 parameters of the
source at the same time:
Voltage Parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot
and Peshoot
Time Parameters: Period, Frequency, Rise Time, Fall Time, + Width, - Width, + Duty
and - Duty
Other Parameters: Area and Period Area.
The all measurement function can measures the parameters of CH1, CH2, CH3 and
CH4 at the same time. You can sent the
:MEASure:AMSource command to set the
source of the all measurement function.
Return
The query returns 0 or 1.
Related
:MEASure:AMSource
:MEASure:AMSource
Syntax :MEASure:AMSource <src>[,<src>[,<src>[,<src>]]]
Set or query the source of the all measurement function.
Parameter
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
Return
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example
:MEASure:AMSource CHANnel1,CHANnel3
Comentarios a estos manuales