Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-7
:ACQuire:SRATe?
Syntax
:ACQuire:SRATe?
Description
Query the current sample rate. The default unit is Sa/s.
Explanation Sample rate is the sample frequency of the oscilloscope, namely the waveform points
sampled per second.
The relationship among the memory depth, sample rate and waveform length is
satisfied with the following equation:
Memory Depth Sample Rate × Waveform Length
Wherein, the Memory Depth can be set using the
:ACQuire:MDEPth command and
the Waveform Length is the product of the horizontal timebase (set by
the
:TIMebase[:MAIN]:SCALe command) and the number of the horizontal scales
(12 for DS1000Z).
Return
Format
The query returns the sample rate in scientific notation.
Example
:ACQuire:SRATe? /*The query returns 2.000000e+09*/
Related
Command
:ACQuire:MDEPth
:TIMebase[:MAIN]:SCALe
:CALibrate Commands
Command List:
:CALibrate:QUIT
:CALibrate:STARt
:CALibrate:QUIT
Syntax
:CALibrate:QUIT
Description
Exit the calibration at any time.
Related
Command
:CALibrate:STARt
:CALibrate:STARt
Syntax
:CALibrate:STARt
Description
The oscilloscope starts to execute self-calibration.
Explanation The self-calibration can make the oscilloscope quickly reach its optimum working
state to obtain the most accurate measurement values.
During the self-calibration, all the channels of the oscilloscope must be disconnected
from the inputs.
The functions of most of the keys are disabled during the self-calibration. You can
sent the :CALibrate:QUIT command to quit the self-calibration.
Related
Command
:CALibrate:QUIT