Rigol DS1000Z Series Especificaciones Pagina 132

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 172
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 131
RIGOL Chapter 2 Command System
2-118 DS1000Z Programming Guide
:TRIGger:NEDGe (Option)
Command List:
:TRIGger:NEDGe:SOURce
:TRIGger:NEDGe:SLOPe
:TRIGger:NEDGe:IDLE
:TRIGger:NEDGe:EDGE
:TRIGger:NEDGe:LEVel
:TRIGger:NEDGe:SOURce
Syntax :TRIGger:NEDGe:SOURce <source>
:TRIGger:NEDGe:SOURce?
Description
Set or query the trigger source in Nth edge trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example :TRIGger:NEDGe:SOURce CHANnel2
:TRIGger:NEDGe:SLOPe
Syntax :TRIGger:NEDGe:SLOPe <slope>
:TRIGger:NEDGe:SLOPe?
Description
Set or query the edge type in Nth edge trigger.
Parameter
Name
Type
Range
Default
<slope> Discrete {POSitive|NEGative} POSitive
Explanation POSitive: trigger on the rising edge of the input signal when the voltage level meets
the specified trigger level.
NEGative: trigger on the falling edge of the input signal when the voltage level meets
the specified trigger level.
Return
Format
The query returns POS or NEG.
Example
:TRIGger:NEDGe:SLOPe NEGative
Vista de pagina 131
1 2 ... 127 128 129 130 131 132 133 134 135 136 137 ... 171 172