RIGOL Chapter 2 Command System
2-2 DS1000Z Programming Guide
:AUToscale
Syntax
:AUToscale
Description Enable the waveform auto setting function. The oscilloscope will automatically adjust the
vertical scale, horizontal timebase and trigger mode according to the input signal to
realize optimum waveform display. This command is equivalent to press the AUTO key at
the front panel.
Explanation Theoretically, waveform auto setting function requires that the frequency of sine can
not be lower than 41Hz, the duty cycle should be greater than 1% as well as the
amplitude must be greater than 20mVpp for square.
When the pass/fail function is enabled (see the
:MASK:ENABle command), if you sent
this command, the oscilloscope will disable the pass/fail function firstly and the
execute waveform auto setting function.
When the waveform record function is enabled or during playback the recorded
waveform, this command is invalid.
:CLEar
Syntax
:CLEar
Description Clear all the waveforms on the screen. If the oscilloscope is in RUN state, waveform will
still be displayed. This command is equivalent to press the CLEAR key at the front panel.
Related
Command
:DISPlay:CLEar
:RUN
:STOP
Syntax :RUN
:STOP
Description The :RUN command makes the oscilloscope starting running and the :STOP command
makes the oscilloscope stopping running. These commands are equivalent to press the
RUN/STOP key at the front panel.
Explanation When the waveform record function is enabled or during playback the recorded
waveform, this command is invalid.