RIGOL
User‟s Guide for DS1000CA Series
2-62
Pass/Fail
The Pass/Fail function monitors changes of signals and output pass or fail signals by
comparing the input signal is within the pre-defined mask. The testing results not
only can be displayed on screen or output by isolated pass/fail port but also can be
alarmed if turn on system sound.
Press UtilityPass/Fail to go to the following menu.
Figure 2-84 Table 2-51 The Pass/Fail menu (Page 1/2)
Menu
Setting
Comments
Enable Test
ON
OFF
Turn on Pass/Fail test.
Turn off Pass/Fail test.
Source
CH1
CH2
Select Pass/Fail test on CH1.
Select Pass/Fail test on CH2.
Operate
(RUN)
(STOP)
Pass/Fail test stopped, press to run.
Pass/Fail test running, press to stop.
Msg display
ON
OFF
Turn on Pass/Fail information display.
Turn off Pass/Fail information display.
Figure 2-85 Table 2-52 The Pass/Fail menu (Page 2/2)
Menu
Settings
Comments
Output
Fail
Fail +
Pass
Pass +
Output when Fail condition detected.
Output and beep when Fail condition
detected.
Output when Pass condition detected.
Output and beep when Pass condition
detected.
Stop
on Output
ON
OFF
Stop test when output occur.
Continue test when output occur.
Mask Setting
Go to mask setting menu.