User‟s Guide for DS1000CA Series
2-28
X-Y Format
This format is useful for studying phase relationships between two signals.
CH1 in the horizontal axis(X) and CH2 in the vertical axis(Y), the oscilloscope uses a
none-trigger acquisition mode, data is displayed as dots.
Press MENUTime BaseX-Y to enter X-Y mode.
Figure 2-33 X-Y display format
The oscilloscope could capture waveform with arbitrary sample rate in Y-T mode
(2GSa/s maximum for single channel, 1GSa/s maximum for double channels). While,
in X-Y mode, the sample rate is up to 500MSa/s. Generally, the sample rate is more
lower, the Lissajous graphics is more better.
The following modes or functions will not work in X-Y format.
Cursor Measurements
Pass/Fail Test
REF or MATH Operations
Delayed Scan Mode
Vector Display Mode
Horizontal knob
Trigger Controls