Rigol DS1000CA series Guía de usuario Pagina 122

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 146
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 121
RIGOL
User‟s Guide for DS1000CA Series
3-6
Example 5: Making Cursor Measurements
There are 20 build-in automatic measurements. They can also be conducted using
cursors to make time and voltage measurements of a waveform quickly.
Measure the Peak Frequency of the First Sinc Waveform
To measure the ringing frequency at the rising edge of a signal, do these steps:
1. Press Cursor key to see the Cursor menu.
2. Press Mode to set Manual mode.
3. Press Type to select X.
4. Turn ( ) knob to place cursor A on the first peak of the wave.
5. Turn ( ) knob to place cursor B on the second peak of the wave.
Figure 3-4 Waveform display
Observe the delta in time and frequency displayed on the screen.