Programming Guide for DM3058
1. :rate:voltage:DC
Command Format:
:rate:voltage:DC?
:rate:voltage:DC{fast|medium|slow}
Function:
The command sets or queries the measuring rate of DC voltage.
Explanations:
:rate:voltage:DC?: Query current measuring rate of DC voltage. DCV function has
to be enabled before execute this command.
:rate:voltage:DC{fast|medium|slow}: Set current measuring rate of DC voltage.
For the range and the explanation of parameters please refer to the following
table:
Table 3-7 The range and explanation of parameters
Parameter
Explanation
Mark
Rate
Refresh rate
fast
Fast measure
F
120 reading/s
50 Hz
medium
Medium measure
M
20 reading/s
20 Hz
slow
Slow measure
S
2.5 reading/s
2.5 Hz
Return Format:
The query returns the parameter related to current measuring rate: fast, medium
or slow.
Example:
Set the measuring rate of DC voltage as Medium.
:rate:voltage:DCmedium
2. :rate:voltage:AC
Command Format:
:rate:voltage:AC?
:rate:voltage:AC{fast|medium|slow}
Function:
The command queries or sets the measuring rate of AC voltage.
Vista de pagina 53