Rigol DM3058 Especificaciones Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 264
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
RIGOL
Programming Guide for DM3058
3-9
1. :measure?
Command Format:
:measure?
Function:
The command queries whether a new data is acquired in current trigger setting.
Return Format:
The query returns true or false. If ―a new data is acquired returns ―ture‖, or else
returns ―false‖.
2. :measure
Command Format:
:measure{AUTO|MANU}
Function:
The command sets measure mode as Auto or Manual.
Example:
Set measure mode as Manual:
:measureMANU
3. :measure:voltage:DC?
Command Format:
:measure:voltage:DC?
Function:
The command queries DC voltage.
Explanation:
The command is invalid for Dual Display.
Return Format:
The query returns current DC voltage in the form of scientific notation, for
Vista de pagina 34