RIGOL
Programming Guide for DM3058
Command Format:
:measure:voltage:DC:filter:state?
:measure:voltage:DC:filter:state{ON|OFF}
Function:
The command queries and sets the switch state of DC filter.
Explanations:
:measure:voltage:DC:filter:state?: Query the switch state of DC filter.
:measure:voltage:DC:filter:state{ON|OFF}: Set the switch state of DC filter.
Return Format:
The query return the switch state of DC filter: ON or OFF that denotes turning DC
filter on or off.
Example:
To turn on DC filter:
:measure:voltage:DC:filter:stateON
8. :measure:voltage:AC?
Command Format:
:measure:voltage:AC?
Function:
The command queries AC voltage.
Explanation:
This command is invalide for Dual Display.
Return Format:
The query returns the current AC voltage in the form of 3.941713e-01, unit is V.
9. :measure:voltage:AC
Command Format: