Rigol DM3058 Especificaciones Pagina 49

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 264
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 48
RIGOL
Programming Guide for DM3058
3-23
Explanations:
The range is the same as the measuring range of AC voltage. See Table 3-2.
Return Format:
The query returns the parameter related to current range: 0, 1, 2, 3 or 4 (see
Table 3-2).
29. :measure:continuity?
Command Format:
:measure:continuity?
Function:
The command quries resistance in the state of short circuit.
Return Format:
The query returns the current resistance in the form of 8.888000e+03, and unit is
Ω.
30. :measure:continuity
Command Format:
:measure:continuity{<range>|MIN|MAX|DEF}
Function:
The command sets the limit resistance in short circurt measurement.
Parameters:
Name
Type
Range of Values
Default
<range>
Consecutive Integer
1 Ω~2000 Ω
10
Explanations:
The range is a consecutive Integer among 1 Ω~2000 Ω.
The default resistance is 10 Ω. MAX denotes the maximum is 2000 Ω. MIN
denotes the minimum is 1 Ω. DEF denotes the default is 10 Ω.