Programming Guide for DM3058
Explanations:
A reading will be force to show on the screen if the instrument has already been
in this mode while you send HOLD.
6. HOLDCLR
Command Format:
HOLDCLR
Function:
The command disable Touch Hold function and back to normal operation.
7. HOLDTHRESH
Command Format:
HOLDTHRESH<threshold>
HOLDTHRESH?
Function:
The command sets or queries the threshold of HOLD.
Explanations:
HOLDTHRESH<threshold>: Set the threshold of HOLD as <threshold>, the
value of it should be 1, 2, 3 or 4 (which separately corresponding to 0.01%, 0.1%,
1%, 10%), any other value will lead an execution error.
HOLDTHRESH?: Query the <threshold>.
Return Format:
The query returns the value of HOLD such as 1, 2, 3 or 4 (0.01%, 0.1%, 1%,
10%).
8. MAX
Command Format:
MAX