Rigol DM3058 Especificaciones Pagina 170

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 264
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 169
RIGOL
Programming Guide for DM3058
Explanations:
[SENSe:]CURRent[:DC]:NULL[:STATe]□{ON|OFF}: Enable or disable DCI NULL
function.
[SENSe:]CURRent[:DC]:NULL[:STATe]?: Query the enable state of DCI NULL
function.
The command ―[SENSe:]CURRent[:DC]:NULL:VALue‖ can be used to set NULL
value.
The commands ―CONFigure:CURRent:DC‖ and ―MEASure:CURRent:DC?‖ can
disable NULL function (OFF).
Return Format:
The query returns the enable state of DCI NULL function.
11. [SENSe:]CURRent[:DC]:NULL:VALue
Command Format:
[SENSe:]CURRent[:DC]:NULL:VALue□{<value>|MIN|MAX}
[SENSe:]CURRent[:DC]:NULL:VALue?□[{MIN|MAX}]
Function:
The command sets or queries the NULL value of DCI measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]CURRent[:DC]:NULL:VALue □ {<value>|MIN|MAX}: Set the NULL
valueof DCI measurement. MIN is -120% max range of current function, MAX is
120% max range of current function, DEF = 0.
[SENSe:]CURRent[:DC]:NULL:VALue?□[{MIN|MAX}]: Query the NULL value of
DCI measurement.
Before sending the command, do run the command ―[SENSe:]CURRent[:DC]:
NULL[:STATe]□ON‖ to enable NULL function.
Return Format:
The query returns the NULL value of DCI measurement.
Vista de pagina 169
1 2 ... 165 166 167 168 169 170 171 172 173 174 175 ... 263 264

Comentarios a estos manuales

Sin comentarios