RIGOL
Programming Guide for DM3058
1. COMP
Command Format
COMP
COMP?
Function:
COMP: Disable the Compare function (COMP). Touch Hold will be actived
automaticlly (you can disable it via HOLDCLR command).
COMP? Query the result of last comparation.
Return Format:
The query returns the result of last comparation. If the measuring result is higher
than the maximum of range, it returns LO; if belong to the range, it returns PASS;
or returns ―—‖ if the measuring has not been finished yet.
2. COMPCLR
Command Format:
COMPCLR
Function:
The command disables the Compare function together with Touch Hold and then
back to normal work. If Touch Hold mode selected, also exit.
3. COMPHI
Command Format:
COMPHI<high value>
Function:
The command sets the upper limit (HI) of COMP as <high value>.
Explanation:
<high value> could be either a signed integer or a signed real number.