Rigol DM3058 Manual de usuario Pagina 58

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 192
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 57
RIGOL
Programming Guide for DM3058/DM3058E
3-33
Table 3-6 DC current range and
the corresponding resolution
Parameter Range Resolution
0 200 μA 1 nA
1 2 mA 10 nA
2 20 mA 100 nA
3 200 mA 1 μA
4 2 A 10 μA
5 10 A 100 μA
MIN 200 μA 1 nA
MAX 10 A 100 μA
DEF 200 mA 1 μA
Example:
Sets the DC current range to the maximum (10A).
:MEASure:CURRent:DC MAX or :MEASure:CURRent:DC 5
13. :MEASure:CURRent:DC:RANGe?
Command Format:
:MEASure:CURRent:DC:RANGe?
Function:
The command queries the DC current range.
Return Format:
The query returns: 0, 1, 2, 3, 4 or 5 (see Table 3-6).
14. :MEASure:CURRent:DC:FILTer[:STATe]
Command Format:
:MEASure:CURRent:DC:FILTer[:STATe]?
:MEASure:CURRent:DC:FILTer[:STATe] {ON|OFF|1|0}
Function:
The commands query and set the AC filter state under the DC current
Vista de pagina 57