PLC.
[SENSe:]FRESistance:NPLC? [MINimum|MAXimum] – Queries the integration
time of 4-wire resistance measurement.
The integration time relates to resolution and measurement rate, refer to Table
4-2.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.
Return Format:
The query returns the current integration time of 4-wire resistance measurement.
28. [SENSe:]FREQuency:APERture
Command Format:
[SENSe:]FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
[SENSe:]FREQuency:APERture? [MINimum|MAXimum]
Function:
The commands set and query the aperture time (or gate time) of frequency
measurement.
Explanations:
[SENSe:]FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum} Sets the
desired aperture time for frequency measurement. The unit is s, MIN=10ms,
MAX=1s, and the default is 100ms.
[SENSe:]FREQuency:APERture? [MINimum|MAXimum] Queries the aperture
time of frequency measurement.
Return Format:
The query returns the current aperture time of frequency measurement.