Rigol DM3058 Manual de usuario Pagina 138

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 192
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 137
RIGOL
Programming Guide for DM3058/DM3058E
4-45
setting of resistance.
Return Format:
The query returns the current precision of 2-wire resistance measurement.
23. [SENSe:]FRESistance:RESolution
Command Format:
[SENSe:]FRESistance:RESolution {<resolution>|MINimum|MAXimum}
[SENSe:]FRESistance:RESolution? [MINimum|MAXimum]
Function:
The commands set and query the precision of 4-wire resistance measurement.
Parameter:
Name Type Range Default
<resolution> Discrete Refer to Table 4-2 3ppm x range
Explanations:
[SENSe:]FRESistance:RESolution {<resolution>|MINimum|MAXimum}– Sets the
desired precision of 4-wire resistance measurement
[SENSe:]FRESistance:RESolution? [MINimum|MAXimum] Queries the current
precision of 4-wire resistance measurement.
<resolution> relates to measurement rate and integration time, refer to Table
4-2.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.
Return Format:
The query returns current precision of 4-wire resistance measurement.
24. [SENSe:]VOLTage:DC:NPLC
Command Format:
[SENSe:]VOLTage:DC:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum}
[SENSe:]VOLTage:DC:NPLC? [MINimum|MAXimum]
Vista de pagina 137