2-4
1. *CLS
Command Format:
*CLS
Function:
The command resets the values in all Event Registers and clears the Error Queue
information.
2. *ESE
Command Format:
*ESE <enable_value>
*ESE?
Function:
The commands set and query the enable value of Event Status Register (ESR).
Explanations:
*ESE <enable_value> - Sets the desired enable value of ESR.
*ESE? - Queries the current enable value of ESR.
<enable_value> - Ranging from 0 to 189.
Return Format:
The query returns the enable value of ESR.
3. *ESR?
Command Format:
*ESR?
Function:
The command queries the event value of Event Status Register (ESR).
Return Format:
The query returns the current event value of ESR.