RIGOL
Programming Guide for DM3058/DM3058E
3-28
The command returns the measured DC voltage value in scientific notation, such
as: 8.492853e-05, the unit is V.
4. :MEASure:VOLTage:DC
Command Format:
:MEASure:VOLTage:DC {<range>|MIN|MAX|DEF}
Function:
The command sets the range and resolution of DC voltage measurement.
Parameter:
Name Type Range Default
<range> Discrete {0|1|2|3|4} 2
Explanations:
The corresponding resultion will be set automatically after you set the range.
The measurement type will change into Manualautomaticlly as you set the
range.
For the parameter range and the corresponding resolution, see table below.
Table 3-4 DC voltage range and
the corresponding resolution
Parameter Range Resolution
0 200 mV 100 nV
1 2 V 1 μV
2 20 V 10 μV
3 200 V 100 μV
4 1000 V 1 mV
MIN 200 mV 100 nV
MAX 1000 V 1 mV
DEF 20 V 10 μV
Example:
Sets the DC votage range to 200V.
:MEASure:VOLTage:DC 3
Sets the DC votage range to the minimum.
Vista de pagina 52