2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-345
:WAVeform:MODE
Syntax
:WAVeform:MODE <mode>
:WAVeform:MODE?
Description
Set or query the data reading mode of the
:WAVeform:DATA? Command.
Parameter
Name Type Range Default
<mode> Discrete {NORMal|RAW|MAXimum} NORMal
Explanation
NORMal: read the waveform data currently displayed on the screen.
RAW: read the waveform data in the internal memory.
Note: The waveform data in the internal memory can only be read when the oscilloscope is in the
stop state and please do not operate the oscilloscope during the reading process.
MAXimum: in the run state, read the waveform data displayed on the screen; in the stop state, read
the waveform data in the internal memory.
Return Format
The query returns NORM, MAX or RAW.