2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-209
:MEASure:MPARea
Syntax 1
:MEASure:MPARea <chan>
Description 1
Enable the period area measurement function of the specified channel.
Syntax 2
:MEASure:MPARea? <chan>
Description 2
Enable the period area measurement function of the specified channel and return the measurement result
(the unit is Vs).
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the period area measurement result in scientific notation. When the measurement
result is invalid, the query returns 9.9e37.
Example
:MEASure:MPARea CHANnel1
:MEASure:MPARea? CHANnel1
You can use the following commands to enable the period area measurement function of the specified
channel and the statistic function of the measurement results as well as query the statistic maximum
value, minimum value, current value, average value and standard deviation of the period area
measurement results.
:MEASure:MPARea:SMAXimum
:MEASure:MPARea:SMINimum
:MEASure:MPARea:SCURrent
:MEASure:MPARea:SAVerage
:MEASure:MPARea:SDEViation