RIGOL 2 Command System
2-114 MSO2000A/DS2000A Programming Guide
:FUNCtion:WANalyze:SETup:THReshold
Syntax
:FUNCtion:WANalyze:SETup:THReshold <NR1>
:FUNCtion:WANalyze:SETup:THReshold?
Description
Set or query the threshold of waveform analysis and the unit is %.
Parameter
Name Type Range Default
<NR1> Integer 1 to 99 10
Explanation
The threshold is used to judge whether the data frame is an error frame. The data frame will be judged as
an error frame if the (relative) difference between it and the template is greater than or equal to the
threshold currently set.
Return Format
The query returns an integer between 1 and 99.
:FUNCtion:WANalyze:SETup:XMASk
:FUNCtion:WANalyze:SETup:YMASk
Syntax
:FUNCtion:WANalyze:SETup:XMASk <x>
:FUNCtion:WANalyze:SETup:XMASk?
:FUNCtion:WANalyze:SETup:YMASk <y>
:FUNCtion:WANalyze:SETup:YMASk?
Description
Set or query the horizontal or vertical adjustment parameter in analysis based on the pass/fail mask and
the unit is div.
Parameter
Name Type Range Default
<x> Real 0.02 to 4, the step is 0.02 0.24
<y> Real 0.04 to 5.12, the step is 0.04 0.48
Return Format
The query returns the horizontal or vertical adjustment parameter in analysis based on the pass/fail mask
in scientific notation.