:MEASure:MARea:SMAXimum
:MEASure:MARea:SMINimum
:MEASure:MARea:SCURrent
:MEASure:MARea:SAVerage
:MEASure:MARea:SDEViation
Syntax 1
:MEASure:MARea:SMAXimum <chan>
:MEASure:MARea:SMINimum <chan>
:MEASure:MARea:SCURrent <chan>
:MEASure:MARea:SAVerage <chan>
:MEASure:MARea:SDEViation <chan>
Description 1
Enable the area measurement function of the specified channel and the statistic function of the
measurement results.
Syntax 2
:MEASure:MARea:SMAXimum? <chan>
:MEASure:MARea:SMINimum? <chan>
:MEASure:MARea:SCURrent? <chan>
:MEASure:MARea:SAVerage? <chan>
:MEASure:MARea:SDEViation? <chan>
Description 2
Query the statistic maximum value, minimum value, current value, average value or standard deviation of
the area measurement results when the area measurement function of the specified channel and the
statistic function of the measurement results are enabled.
Parameter
Name
Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the statistic maximum value, minimum value, current value, average value or standard
deviation of the area measurementresults in scientific notation. When the measurement results are
invalid, the query returns 9.9e37.