Chapter 3 Performance Test RIGOL
Service Guide for DS1000E, DS1000D Series
Specifications Test
Purpose:
Test if all the specifications are in the specified range.
Requirement:
Verify whether the oscilloscope is qualified through FLUKE calibration instrument and
relative softwares.
Tools:
A set of DS1000E, DS1000D series oscilloscope
A set of FLUKE 9500B calibration instrument for oscilloscope and accessories to it
Auxiliary testing softwares
A computer with GPIB and USB interfaces
A GPIB cable
A standard USB cable (Type AB)
Three BNC terminals with 50Ω of resistance
Steps:
1. Connect the FLUKE oscilloscope with PC using a GPIB cable.
Figure 3-14 Connect the testing device
2. Connect the oscilloscope with PC using an USB cable.
3. Use the testing software to test the specifications listed in the table below. Check
if all the specifications are in the specified range.
Table 3-2 Testing specifications and the ranges
Input impedance:
CH1:100mV
500mV
CH2:100mV
500mV
1MΩ±2%, in parallel with
15pF± 3pF
1MΩ±2%, in parallel with
15pF± 3pF
PROBE (9500B accessories)