RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-8
P/F Interface Test
Purpose:
Test if P/F interface work normally through RS5101 module.
Tools:
A set of DS1000E, DS1000D series digital oscilloscope
A BNC cable
A RD5101 module
Steps:
1. Power on the oscilloscope and RD5101 respectively.
2. Connect P/F interface on DS1000E, DS1000D with [Pass/Fail out] interface on
RD5101 by BNC cable.
Figure 3-10 P/F interface connection
3. Turn on the oscilloscope, press Storage Storage Factory to recall the
factory settings.
4. Press UtilityPass/fail to enter the setting interface of Pass/Fail, see the table
below
Table 3-1 Pass/Fail setting items
Items
Setting
Enable Test
ON
Source
CH1
Msg display
ON
Output
Fail
Stop on Output
OFF
5. After setting, press Operate to start Pass/Fail test.
6. Observe the reminder light of PASS/FAIL INDICATO of RRD5101; if the reminder
light twinkles at a frequency, which indicates that the interface works normally, if
no response to the lighter, errors might happened to the interface.