Chapter 1 Specifications RIGOL
Service Guide for DS1000E, DS1000D Series
1-3
A/D converter
8-bit resolution, each channel samples simultaneously
[2]
Volts/div Range
2mV/div~10V/div at input BNC
Maximum Input
Analog channel maximum input voltage
CAT I 300Vrms, 1000Vpk; instantaneous voltage 1000Vpk
CAT II 100Vrms, 1000Vpk
RP2200 10:1: CAT II 300Vrms
RP3200 10:1: CAT II 300Vrms
RP3300 10:1: CAT II 300Vrms
Offset Range
± 40V(250mV/div~10V/div),± 2V(2mV/div~245mV/div)
Analog Bandwidth
100MHz (DS1102D,DS1102E)
50MHz (DS1052D, DS1052E)
Single-shot
Bandwidth
100MHz (DS1102D, DS1102E)
50MHz (DS1052D, DS1052E)
Selectable Analog
Bandwidth Limit
(typical)
20MHz
Lower Frequency
Limit (AC 3dB)
5Hz (at input BNC)
Rise Time at BNC,
typical
<3.5ns, <7ns,
On (100MHz, 50MHz) respectively
Dynamic range
± 5div
DC Gain Accuracy
2mV/div-5mV/div:
± 4% (Normal or Average acquisition mode)
10mV/div-10V/div:
± 3% (Normal or Average acquisition mode)
DC Measurement
Accuracy, Average
Acquisition Mode
Average of ≥16 Waveforms with vertical position at zero:
±(DC Gain Accuracy× reading+0.1div+1mV)
Average of ≥16 Waveforms with vertical position not at zero:
±[DC Gain Accuracy× (reading+vertical position)+(1% of
vertical position) + 0.2div]
Add 2mV for settings from 2mV/div to 245 mV/div
Add 50mV for settings from >250mV/div to 10V/div
Delta Volts
Measurement
Accuracy (Average
Acquisition Mode)
Delta Volts between any two averages of 16 waveforms
acquired under same setup and ambient
conditions: ± (DC Gain Accuracy× reading + 0.05 div)
Trigger
Trigger Sensitivity
0.1div~1.0div (adjustable)
Trigger Level Range
Internal
± 6 divisions from center of screen
EXT
± 1.2V
Trigger Level Accuracy
(typical) applicable for
the signal of rising and
falling time ≥20ns
Internal
± (0.3div × V/div)(± 4 divisions from center of
screen)
EXT
± (6% of setting + 200 mV)