RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-44
The query returns PASS or FAIL.
Explanation Limit operation function must be enabled before sending this
command.
17. :CALCulate:LIMIt:LOWEr
Syntax :CALCulate:LIMIt:LOWEr?
:CALCulate:LIMIt:LOWEr {
<range>
|MIN|DEF}
Function Set the lower limit of Limit operation.
The query returns the lower limit of Limit operation under the current
measurement function in scientific notation.
Explanation The lower limit should not be greater than the upper limit. For
more details, refer to :CALCulate:LIMIt:UPPEr.
The range of
<range>
depends on the current measurement
function.
Measurement <range> Default Unit
DC voltage ±1200 0 V
AC voltage 0 - 900 0 V
DC current ±12 0 A
AC current 0 - 12 0 A
Resistance 0 - 1.2e+08 0 Ω
Capacitance 0 - 2.4e-04 0 F
Frequency 3 - 3.0e+05 3 Hz
Period 3.0e-06 - 3.0e-01 3.0e-06 s
Ratio ±1.0e+09 -1.0e+09
18. :CALCulate:LIMIt:UPPEr
Syntax :CALCulate:LIMIt:UPPEr?
:CALCulate:LIMIt:UPPEr {
<range>
|MAX|DEF}
Function Set the upper limit of Limit operation.
The query returns the upper limit of Limit operation under the current
measurement function in scientific notation.
Explanation The upper limit should not be lower than the lower limit. For more
details, refer to :CALCulate:LIMIt:LOWEr.
The range of
<range>
depends on the current measurement
function.
Measurement Range Default Unit
DC voltage ±1200 1 V