Chapter 2 Command System RIGOL
Programming Guide for DM3000
2-3
1. *CLS
Syntax
*CLS
Function
Clear all the event registers and the error queue.
2.
*IDN?
Syntax
*IDN?
Function The query returns the instrument ID character string. For example,
Rigol Technologies,DM3064,DM3A083100011,03.12.00.03.09.00
3.
*RST
Syntax *RST
Function
Reset the instrument and restore it to factory defaults.
4. CMDSet
Syntax CMDSet?
CMDSet {RIGOL|AGILENT|FLUKE}
Function Specify the command set currently used by the instrument.
The query returns RIGOL, AGILENT or FLUKE.
Default RIGOL
Note: unless otherwise noted, the return values in this manual do not