RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-42
Syntax :CALCulate:NULL:OFFSet?
:CALCulate:NULL:OFFSet {
<range>
|MIN|MAX|DEF}
Function Set the offset of Null operation.
The query returns the offset of Null operation under the current
measurement function in scientific notation.
Explanation The offset value can hold seven digits after the decimal point.
The range of
<range>
is different for different measurement
function.
Measurement <range> Default Unit
DC voltage ±1200 0 V
AC voltage ±900 0 V
DC current ±12 0 A
AC current ±12 0 A
Resistance ±1.2e+08 0 Ω
Capacitance ±2.4e-04 0 F
Frequency 3 - 3.0e+05 0 Hz
Example Set the offset of Null operation under DC voltage measurement to
10.2010031V:
:CALCulate:NULL:OFFSet 10.2010031
The query returns: +1.02010031e+01.
9. :CALCulate:DB:STATe
Syntax :CALCulate:DB:STATe?
:CALCulate:DB:STATe {ON|OFF|1|0}
Function Enable or disable the dB operation function.
The query returns ON or OFF.
Default OFF
10.
:CALCulate:DB?
Syntax
:CALCulate:DB?
Function The query returns the dB measurement value in scientific notation, for
example, -4.14956621e+01.
Explanation dB operation function must be enabled before sending this command.
11.
:CALCulate:DB:REFErence
Syntax :CALCulate:DB:REFErence?
:CALCulate:DB:REFErence {
<range>
|MIN|MAX|DEF}