Rigol DM3058 Manual de usuario Pagina 50

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 138
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 49
RIGOL
Users Guide for DM3058/DM3058E
2-28
3. Set the relative value (Optional)
Press REL to open or close Relative operation function. When it is open, “REL
is shown over the screen and the reading displayed is a relative value which
comes from the result of Actual measurement value subtracts the value that
has been set. (For the details about how to set, please refer to “
Math
Functions” in Chapter 2.)
4. Read the measurement value
Period measurement is fixed at “Slow” rate, 5.5 digits resolution. Therefore,
you cannot adjust the reading rate when reading the result.
5. View history data
Press History to enter into the following interface and check the data getting
from this measurement and save.
Figure 2-35
View Interface of History Information
There are three types of way for viewing historical date: “Info”,List” and
HistoG”. Press Save after checking. By pressing Update, the History
information will be refreshed to be current newest information.
Vista de pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 137 138

Comentarios a estos manuales

Sin comentarios