To Measure Period
Operating Steps:
1. Press
until enter into the following interface.
Figure 2-33 Period Measurement Interface
2. Connect the red lead to terminal-HI and black lead to terminal LO as the figure
below.
Figure 2-34 Sketch map for connecting
Table 2-11 Period Characteristic
Ranges 200mV, 2V, 20V, 200V, 750V
Measurement Range
1μs0.05s
Input Protection 750VRMS (HI)
Configurable Parameters REL
AC Signal