10 Waveform Record RIGOL
DS2000 User’s Guide
10-3
3. Interval
Set the time interval between the frames in waveform record and the range
available is from 100 ns to 10 s.
4. Total Frames
The menu shows the maximum number of frames that can be recorded
currently.
As the capacity of the waveform memory is fixed, the more the number of points
each frame of waveform has, the less the number of waveform frames can be
recorded. Thus, the maximum end frame of waveform record is decided by the
“Memory Depth” currently selected. Please refer to the instruction in Memory
Depth” to select the desired memory depth.
Table 10-1 Memory Depth and Total Frames
Memory Depth Maximum End Frame
Auto 65000
14 k points 8128
140 k points 508
1.4M points 63
14M points 7