RIGOL 6 To Make Measurements
DS2000 User’s Guide
6-18
Voltage Parameters
1. Vmax: the voltage value from the highest point of the waveform to the GND.
2. Vmin: the voltage value from the lowest point of the waveform to the GND.
3. Vpp: the voltage value from the highest point to the lowest point of the
waveform.
4. Vtop: the voltage value from the flat top of the waveform to the GND.
5. Vbase: the voltage value from the flat base of the waveform to the GND.
6. Vamp: the voltage value from the top of the waveform to the base of the
waveform.
7. Vavg: the arithmetic average value on the whole waveform or on the gating
area.
n
x
Average
i
=
, wherein,
i
x
is the ith point being measured,
n
is the number
of points being measured.
8. Vrms: the root mean square value on the whole waveform or the gating area.
n
x
RMS
n
1i
2
i
=
=
, where,
i
x
is the ith point being measured,
n
is the number of
points being measured.
9. Overshoot: the ratio of the difference of the maximum value and top value of
Preshoot
Overshoot
Vamp
Vbase
Vmax
Vmin
Vpp
Vtop