RIGOL User’s Guide DS2000 Series Digital Oscilloscope Sept. 2012 RIGOL Technologies, Inc.
RIGOL DS2000 User’s Guide VIII Pollution Degree Degree 2 Pollution Degree Definitions Pollution degree 1: No pollution or only dry, non-conductive po
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-24 PAL: the frame frequency is 25 frames per second. The TV sweep line is 625 with the o
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-25 Pattern Trigger Identify a trigger condition by looking for a specified pattern.
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-26 or : set the pattern to the rising or falling edge of the channel selected. Note
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-27 Delay Trigger (Option) Trigger when the time difference (△T) between the specifie
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-28 Delay Type: Press DelayType to set the time limit condition of delay trigger. >: t
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-29 TimeOut Trigger (Option) Trigger when the time interval (△T) from when the rising
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-30 Timeout Time: Press TimeOut to set the timeout time of timeout trigger and the range is
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-31 Duration Trigger (Option) Identify a trigger condition by looking for the duratio
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-32 Trigger Condition: Press When to select the desired trigger condition. >: trigger
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-33 Setup/Hold Trigger Trigger when the internal state of the setup or hold time rela
RIGOL DS2000 User’s Guide IX General Care and Cleaning General Care: Do not store or leave the instrument at places where the instrument will be
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-34 Setup Type: Press SetupType to select the desired setup type. Setup: set the time tha
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-35 RS232 Trigger Trigger according to the start frame, error frame, check error or d
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-36 --press Data Bits to select “5 bit”, “6 bit”, “7 bit” or “8 bit”; --press Data and inpu
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-37 I2C Trigger Trigger on the start condition, restart, stop, missing acknowledgemen
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-38 behind the preset address (Write, Read or R/W direction). After this trigger condition i
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-39 Trigger Setting: Press Setting to set the trigger parameter (noise rejection) unde
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-40 SPI Trigger Trigger on the data pattern on the specified edge. When using SPI trigger
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-41 can be set to any integer between 4 and 32. Press CurrentBit to set the number
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-42 USB Trigger (Option) Trigger on the SOP, EOP, RC, Suspend and Exit Suspend of the dat
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-43 normal or single. The corresponding status light of the current trigger mode turns
RIGOL DS2000 User’s Guide X Environmental Considerations The following symbol indicates that this product complies with the applicable European Union
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-44 Trigger Output Connector The trigger output connector at the rear panel can output trig
6 To Make Measurements RIGOL DS2000 User’s Guide 6-1 6 To Make Measurements DS2000 can make math operation, cursor measurement and auto measur
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-2 Math Operation DS2000 can realize various math operations (including: addition (A+B), subtracti
6 To Make Measurements RIGOL DS2000 User’s Guide 6-3 Substraction Subtract the waveform voltage values of signal source B from that of source
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-4 Division Divide the waveform voltage values of signal source A by that of source B point by poi
6 To Make Measurements RIGOL DS2000 User’s Guide 6-5 FFT FFT is used to quickly perform Fourier transform on specified signals and transform t
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-6 characteristics. Press Window to select the desired window function and the default is “Rectangl
6 To Make Measurements RIGOL DS2000 User’s Guide 6-7 4. Set the Vertical Scale In FFT measurement, the unit of the horizontal axis changes fro
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-8 Logic Operation Perform logic operation on the waveform voltage values of the specified sources
6 To Make Measurements RIGOL DS2000 User’s Guide 6-9 Press MATH Operate to select “Logic”: Press Log.Formula to select the desired operati
RIGOL DS2000 User’s Guide XI DS2000 Series Overview DS2000 is a high-performance and low bandwidth digital oscilloscope developed on the basis of
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-10 Advanced Operation DS2000 provides advanced operation function that allows users to define ope
6 To Make Measurements RIGOL DS2000 User’s Guide 6-11 2. Channel You can select any channel (CH1 or CH2). 3. Function Please refer to the f
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-12 with 10 as the bottom number in the variables. The range is from -9 to 9. For example, Variable
6 To Make Measurements RIGOL DS2000 User’s Guide 6-13 Auto Measurement DS2000 provides auto measurements of 24 waveform parameters and the sta
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-14 this softkey is pressed, the quick parameter measurement items might not be displayed in the po
6 To Make Measurements RIGOL DS2000 User’s Guide 6-15 One-key Measurement of 24 Parameters Press MENU at the left of the screen to turn on the
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-16 Time Parameters 1. Period: defined as the time between the middle threshold poin
6 To Make Measurements RIGOL DS2000 User’s Guide 6-17 Delay and Phase Source ASource BPeriodDelay Source A and source B can be CH1 or CH2.
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-18 Voltage Parameters 1. Vmax: the voltage value from the highest point o
6 To Make Measurements RIGOL DS2000 User’s Guide 6-19 the waveform to the amplitude value. 10. Preshoot: the ratio of the difference of the min
RIGOL DS2000 User’s Guide XII Conform to LXI-C instrument standards. Enable quick, economic and efficient creation and reconfiguration of test syst
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-20 Other Parameters 1. Area: the area of the whole waveform within the screen and the unit is
6 To Make Measurements RIGOL DS2000 User’s Guide 6-21 Frequency Counter Measurement The hardware frequency counter supplied with this oscill
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-22 Measurement Setting 1. Source Selection Press Measure Source to select the desired channel
6 To Make Measurements RIGOL DS2000 User’s Guide 6-23 4. Phase Measurement Setting Specify the source A and source B in the measurement items
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-24 To Clear the Measurement If you have currently enabled one or more items in the 24 measurement
6 To Make Measurements RIGOL DS2000 User’s Guide 6-25 All Measurement All measurement could measure all the time and voltage parameters (each
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-26 Statistic Function Make statistic and display the current, average, minimum (or standard devia
6 To Make Measurements RIGOL DS2000 User’s Guide 6-27 Measurement History To view the history measurement data, press Measure MeasHistory M
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-28 Cursor Measurement Cursors are the horizontal and vertical marks that can be used to measure t
6 To Make Measurements RIGOL DS2000 User’s Guide 6-29 Manual Mode In this mode, a pair of cursors will appear. You can adjust the cursors manu
RIGOL DS2000 User’s Guide XIII Document Overview 1 Quick Start Provide information about preparations before using the instrument and a brief int
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-30 3. Select X (Y) Axis Unit When the cursor type is “Time”, press Time Unit to select “s”, “Hz”,
6 To Make Measurements RIGOL DS2000 User’s Guide 6-31 adjust the position of cursor A and B at the same time. During the adjustment, the measur
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-32 Track Mode In this mode, one or two pairs of cursors will appear. You can adjust the two pairs
6 To Make Measurements RIGOL DS2000 User’s Guide 6-33 If needed, please refer to the following steps to modify manual cursor track measurement
RIGOL 6 To Make Measurements DS2000 User’s Guide 6-34 Cursor Track (Before Horizontal Expansion): Cursor Track (After Horizontal Expansion):
6 To Make Measurements RIGOL DS2000 User’s Guide 6-35 Auto Mode In this mode, one or more cursors will appear. You can use auto cursor measure
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-1 7 Protocol Decoding Protocol analysis can be used by users to discover errors, debug hardwar
RIGOL 7 Protocol Decoding DS2000 User’s Guide 7-2 Parallel Decoding Parallel bus consists of clock line and data line. As shown in the figure below,
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-3 example, when the bus bits is 20, the range available is 0, 1…19. Next, press Channel to spec
RIGOL DS2000 User’s Guide XIV Format Conventions in this Manual: Front panel key: denoted by the format of “Text Box + Button Name (Bold)”, for exampl
RIGOL 7 Protocol Decoding DS2000 User’s Guide 7-4 5. Decoding Table The decoding table displays the decoded data and the corresponding line number a
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-5 RS232 Decoding (Option) RS232 serial bus consists of the transmitting data line (TX) and the
RIGOL 7 Protocol Decoding DS2000 User’s Guide 7-6 Start BitData BitCheck BitStop Bit Start Bit: represent when the data starts outputting. Setting th
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-7 5. Data Packet Setting As mentioned before, in RS232, you need to set the start bit, data bi
RIGOL 7 Protocol Decoding DS2000 User’s Guide 7-8 8. The Error Expression during Decoding DS2000 makes full use of the resources such as color and
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-9 Check Error: When check bit error is detected during the decoding, red error mark will be dis
RIGOL 7 Protocol Decoding DS2000 User’s Guide 7-10 I2C Decoding (Option) I2C serial bus consists of the clock line (SCLK) and the data line (SDA).
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-11 Press BusStatus to turn the bus display on or off. 4. Decoding Table The decoding table di
RIGOL 7 Protocol Decoding DS2000 User’s Guide 7-12 When the ACK (ACKnowledge Character) is not met, the red error marks as shown in the figure belo
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-13 SPI Decoding (Option) SPI serial bus consists of clock line (SCLK) and data line (SDA). Ho
RIGOL DS2000 User’s Guide XV Contents Guaranty and Declaration ... I Safet
RIGOL 7 Protocol Decoding DS2000 User’s Guide 7-14 4. Endian Setting Press Endian to select “LSB” or “MSB” and the default is “MSB”. 5. Display-re
7 Protocol Decoding RIGOL DS2000 User’s Guide 7-15 Not enough for 7 bits
8 Reference Waveform RIGOL DS2000 User’s Guide 8-1 8 Reference Waveform In actual testing process, the waveform being tested can be compared w
RIGOL 8 Reference Waveform DS2000 User’s Guide 8-2 To Enable REF Function Press REF in the vertical control area (VERTICAL) at the front panel to en
8 Reference Waveform RIGOL DS2000 User’s Guide 8-3 To Set the Color DS2000 series oscilloscope provides five colors (gray, green, light blue,
RIGOL 8 Reference Waveform DS2000 User’s Guide 8-4 To Adjust REF Waveform Display To adjust the reference waveform specified in Current: Press REF
9 Pass/Fail Test RIGOL DS2000 User’s Guide 9-1 9 Pass/Fail Test Monitor the change of the signal by judging whether the input signal is within
RIGOL 9 Pass/Fail Test DS2000 User’s Guide 9-2 To Enable Pass/Fail Test Press Utility Pass/Fail Enable Test to select “ON”. Note that when the t
9 Pass/Fail Test RIGOL DS2000 User’s Guide 9-3 the two cursor lines respectively. Or, press CursorAB and use to adjust the positions of curso
RIGOL DS2000 User’s Guide XVI Navigation Knob ... 1-18 MENU ...
RIGOL 9 Pass/Fail Test DS2000 User’s Guide 9-4 the screen (if display is turned on) and only one pulse is output from the [Trigger Out] connector (if
9 Pass/Fail Test RIGOL DS2000 User’s Guide 9-5 To Load the Test Mask Users can also load the test mask files (*.pf) stored in the internal Fla
10 Waveform Record RIGOL DS2000 User’s Guide 10-1 10 Waveform Record Waveform record can record the waveforms of the input channels (CH1 and C
RIGOL 10 Waveform Record DS2000 User’s Guide 10-2 Waveform Record Waveforms from both the channels currently turned on will be recorded during wavef
10 Waveform Record RIGOL DS2000 User’s Guide 10-3 3. Interval Set the time interval between the frames in waveform record and the range availa
RIGOL 10 Waveform Record DS2000 User’s Guide 10-4 Record Constant On DS2000 provides record “Open” (record constant on) mode for waveform record. Wh
10 Waveform Record RIGOL DS2000 User’s Guide 10-5 2. Enable record constant on mode. 3. Press RUN/STOP to stop the record. 4. Analyze the
RIGOL 10 Waveform Record DS2000 User’s Guide 10-6 Waveform Playback Waveform playback can play back the waveforms currently recorded. Press Utility
10 Waveform Record RIGOL DS2000 User’s Guide 10-7 5. End Frame Press this softkey and use to set the end frame of the playback. The default i
RIGOL DS2000 User’s Guide XVII 5 To Trigger the Oscilloscope ... 5-1 Trigger Source.
RIGOL 10 Waveform Record DS2000 User’s Guide 10-8 Waveform Analysis This function is used to analyze the recorded waveform. Press Utility Record
10 Waveform Record RIGOL DS2000 User’s Guide 10-9 the progress bar is displayed and the parameters can not be modified. After the analysis fini
RIGOL 10 Waveform Record DS2000 User’s Guide 10-10 4. Cancel During the analysis, users can press Cancel to stop the analysis and press Start agai
10 Waveform Record RIGOL DS2000 User’s Guide 10-11 9. Time Tag The time tag is used to display the absolute recording time of each frame of th
RIGOL 10 Waveform Record DS2000 User’s Guide 10-12 Analysis Based on Trace Press Analyze and select “Trace”. Then, set the templet used in analysis
10 Waveform Record RIGOL DS2000 User’s Guide 10-13 Analysis Based on Pass/Fail Mask Press Analyze and select “Pass/Fail”. Then, set the temple
RIGOL 10 Waveform Record DS2000 User’s Guide 10-14 4. Create Mask Mask is the templet used in waveform analysis. Press Create Mask to immediately ap
11 Display Control RIGOL DS2000 User’s Guide 11-1 11 Display Control You can set the type, persistence time and brightness of waveform display
RIGOL 11 Display Control DS2000 User’s Guide 11-2 To Select the Display Type Press Display Type to set the waveform display mode to “Vectors” or
11 Display Control RIGOL DS2000 User’s Guide 11-3 To Set the Persistence Time Press Display Persis.Time to set the persistence time of the o
RIGOL DS2000 User’s Guide XVIII Measurement Setting ... 6-22 To Clear the Measureme
RIGOL 11 Display Control DS2000 User’s Guide 11-4 3. Infinite In this mode, the oscilloscope displays the waveform newly acquired without cleari
11 Display Control RIGOL DS2000 User’s Guide 11-5 To Set the Waveform Intensity Press Display WaveIntensity or turn when the menu is hidde
12 Store and Recall RIGOL DS2000 User’s Guide 12-1 12 Store and Recall Users can save the current settings, waveforms, and screen image of the
RIGOL 12 Store and Recall DS2000 User’s Guide 12-2 Storage System Press Storage to enter the store and recall setting interface. This oscillosco
12 Store and Recall RIGOL DS2000 User’s Guide 12-3 Storage Type Press Storage Storage to select the desired storage type. The default is “Tr
RIGOL 12 Store and Recall DS2000 User’s Guide 12-4 5. CSV Save the waveform data displayed on the screen or of the specified channels in external me
12 Store and Recall RIGOL DS2000 User’s Guide 12-5 Internal Storage and Recall Internal storage and recall support “Setups” in Storage. In the
RIGOL 12 Store and Recall DS2000 User’s Guide 12-6 Note: in internal storage, New File and New Folder are not available. 2. Load the specified typ
12 Store and Recall RIGOL DS2000 User’s Guide 12-7 External Storage and Recall Before using external storage and recall, make sure that the US
RIGOL DS2000 User’s Guide XIX Analysis Based on Pass/Fail Mask ... 10-13 11 Display Control ...
RIGOL 12 Store and Recall DS2000 User’s Guide 12-8 New File or Folder” to create a new file name. Figure g 5) Press OK to execute the saving opera
12 Store and Recall RIGOL DS2000 User’s Guide 12-9 2) As shown in Figure i, use to select the desired file to load and then press Load to lo
RIGOL 12 Store and Recall DS2000 User’s Guide 12-10 Disk Management Press Storage Disk.Manage to turn on the disk management interface as shown in
12 Store and Recall RIGOL DS2000 User’s Guide 12-11 To Select File Type Except the file types in Storage, the oscilloscope can also display, s
RIGOL 12 Store and Recall DS2000 User’s Guide 12-12 To Create a New File or Folder This operation is only valid in external storage. Before using ex
12 Store and Recall RIGOL DS2000 User’s Guide 12-13 English Input Method For example, create a file or folder with the name “Filename”. 1. P
RIGOL 12 Store and Recall DS2000 User’s Guide 12-14 Chinese Input Method For example, create a file or folder with the name “文件名”. 1. Press Keyboa
12 Store and Recall RIGOL DS2000 User’s Guide 12-15 To Delete a File or Folder Folder operation is valid only in external storage. Before usin
RIGOL 12 Store and Recall DS2000 User’s Guide 12-16 To Rename a File or Folder Rename operation is valid only in external storage. Before using the
12 Store and Recall RIGOL DS2000 User’s Guide 12-17 Factory Press Storage Default to return the oscilloscope to its factory state (refer to t
RIGOL DS2000 User’s Guide XX 14 Remote Control ... 14-1 Remote Control via
RIGOL 12 Store and Recall DS2000 User’s Guide 12-18 Amplitude Scale Coarse Channel Unit V Acquisition Setting (Acquire) Acquisition Mode Normal S
12 Store and Recall RIGOL DS2000 User’s Guide 12-19 Cursor B CH1 CurA -4*1 μs CurB 4*1 μs Storage Setting (Storage) Storage Type Traces U
RIGOL 12 Store and Recall DS2000 User’s Guide 12-20 Analyze Analysis Mode Trace Source CH1 Trace Mode Current Frame Current Frame 1 Templet Displ
12 Store and Recall RIGOL DS2000 User’s Guide 12-21 Source CH1 Window Function Rectangle Display Split Scale dB Vertical Scale 20 dB Horiz
RIGOL 12 Store and Recall DS2000 User’s Guide 12-22 Endian LSB Baud 9600 bps Data Bits 8 Stop Bit 1 Even-Odd Check None Packet OFF Packet End
13 System Function Setting RIGOL DS2000 User’s Guide 13-1 13 System Function Setting The contents of this chapter: Remote Interface Config
RIGOL 13 System Function Setting DS2000 User’s Guide 13-2 Remote Interface Configuration DS2000 can communicate with PC via LAN, USB and GPIB (wit
13 System Function Setting RIGOL DS2000 User’s Guide 13-3 IP Configuration Type (DHCP) The configuration type of the IP address can be DHCP, a
RIGOL 13 System Function Setting DS2000 User’s Guide 13-4 1. Set the IP Address The format of IP address is nnn.nnn.nnn.nnn; wherein, the range of
13 System Function Setting RIGOL DS2000 User’s Guide 13-5 Set the Domain Name Server You can set this paramter In Auto IP and Static IP mode.
1 Quick Start RIGOL DS2000 User’s Guide 1-1 1 Quick Start This chapter introduces the preparations when using the oscilloscope for the first t
RIGOL 13 System Function Setting DS2000 User’s Guide 13-6 USB Device This oscilloscope can communicate with PC or PictBridge printer via the USB
13 System Function Setting RIGOL DS2000 User’s Guide 13-7 System-related Sound When the sound is enabled, you can hear the sound of the beepe
RIGOL 13 System Function Setting DS2000 User’s Guide 13-8 System Information Press Utility System System Info to view the version information
13 System Function Setting RIGOL DS2000 User’s Guide 13-9 System Time The system time is displayed at the lower right corner of the screen in
RIGOL 13 System Function Setting DS2000 User’s Guide 13-10 Screen When the oscilloscope enters the idle state and holds for a certain period of ti
13 System Function Setting RIGOL DS2000 User’s Guide 13-11 Self-calibration The self-calibration program can quickly make the oscilloscope rea
RIGOL 13 System Function Setting DS2000 User’s Guide 13-12 Aux Output Users can set the type of the signal output from the [Trigger Out] connect
13 System Function Setting RIGOL DS2000 User’s Guide 13-13 Option Management This oscilloscope provides multiple options to fulfill your measu
14 Remote Control RIGOL DS2000 User’s Guide 14-1 14 Remote Control The oscilloscope can be controlled remotely mainly through the following tw
RIGOL 1 Quick Start DS2000 User’s Guide 1-2 General Inspection 1. Inspect the shipping container for damage. Keep the damaged shipping container or
RIGOL 14 Remote Control DS2000 User’s Guide 14-2 Remote Control via USB 1. Connect the device Connect the oscilloscope (USB DEVICE) with your PC
14 Remote Control RIGOL DS2000 User’s Guide 14-3 5 6 3 4
RIGOL 14 Remote Control DS2000 User’s Guide 14-4 3. Search device resource Start up the Ultra Sigma and the software will automatically search
14 Remote Control RIGOL DS2000 User’s Guide 14-5 4. View the device resource The resources found will appear under the “RIGOL Online Resource”
RIGOL 14 Remote Control DS2000 User’s Guide 14-6 Remote Control via LAN 1. Connect the device Connect the oscilloscope to your LAN using a networ
14 Remote Control RIGOL DS2000 User’s Guide 14-7 Figure (b) Figure (c) 4. View device resource The resources found will appear under the “RIG
RIGOL 14 Remote Control DS2000 User’s Guide 14-8 “SCPI Panel Control” to turn on the remote command control panel (as shown in the figure below) th
14 Remote Control RIGOL DS2000 User’s Guide 14-9 Remote Control via GPIB 1. Connect the device Use the USB to GPIB interface converter to ext
RIGOL 14 Remote Control DS2000 User’s Guide 14-10 5. View device resource Click “OK” to return back to the main interface of Ultra Sigma. The reso
15 Troubleshooting RIGOL DS2000 User’s Guide 15-1 15 Troubleshooting The commonly encountered failures and their solutions are listed below. W
1 Quick Start RIGOL DS2000 User’s Guide 1-3 Appearance and Dimensions Figure 1-1 Front View Unit: mm Figure 1-
RIGOL 15 Troubleshooting DS2000 User’s Guide 15-2 signal should use “Video” trigger. Only when the proper trigger type is used, can the waveform be
16 Specifications RIGOL DS2000 User’s Guide 16-1 16 Specifications All the specifications are guaranteed except the parameters marked with “Ty
RIGOL 16 Specifications DS2000 User’s Guide 16-2 Transient Overvoltage 1000 Vpk with RP2200 10:1 probe: CAT II 300 Vrms with RP3300 10:1 probe: CAT I
16 Specifications RIGOL DS2000 User’s Guide 16-3 DS2102: 20 MHz DS2072: 20 MHz Low Frequency Response (AC Coupling, -3dB) ≤5 Hz (on BNC) Calcul
RIGOL 16 Specifications DS2000 User’s Guide 16-4 Condition Pulse Polarity Positive, Negative Pulse Width Range 2 ns to 4 s Windows Trigger (Option
16 Specifications RIGOL DS2000 User’s Guide 16-5 Data Type H, L Setup Time 2 ns to 1 s Hold Time 2 ns to 1 s RS232/UART Trigger Polarity No
RIGOL 16 Specifications DS2000 User’s Guide 16-6 Period, Rise Time, Fall Time, Positive Pulse Width, Negative Pulse Width, Positive Duty Cycle, Negat
16 Specifications RIGOL DS2000 User’s Guide 16-7 Display Color 160,000 Color (TFT) Persistence Time Min, 50ms, 100ms, 200ms, 500ms, 1 s, 2 s,
RIGOL 16 Specifications DS2000 User’s Guide 16-8 Calibration Interval The recommended calibration interval is one year. Regulatory Information Electr
17 Appendix RIGOL DS2000 User’s Guide 17-1 17 Appendix Appendix A: Accessories and Options Description Order Number Model DS2202 (200 MHz, d
RIGOL 1 Quick Start DS2000 User’s Guide 1-4 To Prepare the Oscilloscope for Use To Adjust the Supporting Legs Adjust the supporting legs properly to
RIGOL 17 Appendix DS2000 User’s Guide 17-2 Appendix B: Warranty RIGOL warrants that its products mainframe and accessories will be free from defects
17 Appendix RIGOL DS2000 User’s Guide 17-3 Appendix C: Any Question or Comment? If you have any question or comment on our document, please ma
Index RIGOL DS2000 User’s Guide 1 Index - Duty ... 6-16 - Width ... 6-16 +
RIGOL Index DS2000 User’s Guide 2 Rectangle ... 6-6 Rise Time ...6-16 Roll ...
1 Quick Start RIGOL DS2000 User’s Guide 1-5 To Connect to Power Supply The power requirements of DS2000 are 100-240 V, 45-440 Hz. Please use t
RIGOL 1 Quick Start DS2000 User’s Guide 1-6 Power-on Inspection When the oscilloscope is energized, press the power key at the lower-left corner o
1 Quick Start RIGOL DS2000 User’s Guide 1-7 To Connect the Probe RIGOL provides passive probes for the DS2000 series oscilloscopes. For detail
RIGOL DS2000 User’s Guide I Guaranty and Declaration Copyright © 2012 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL
RIGOL 1 Quick Start DS2000 User’s Guide 1-8 Function Inspection 1. Press Storage Default to restore the instrument to its default configuration.
1 Quick Start RIGOL DS2000 User’s Guide 1-9 Tip The signal output from the probe compensation connector can only be used for probe compensati
RIGOL 1 Quick Start DS2000 User’s Guide 1-10 Probe Compensation When the probes are used for the first time, you should compensate the probes to mat
1 Quick Start RIGOL DS2000 User’s Guide 1-11 Front Panel Overview Figure 1-9 Front Panel Overview Table 1-1 Front Panel Description No.
RIGOL 1 Quick Start DS2000 User’s Guide 1-12 Rear Panel Overview Figure 1-10 Rear Panel Overview 1. Handle Pull up the handle vertically for
1 Quick Start RIGOL DS2000 User’s Guide 1-13 oscilloscope at each trigger via this interface. 5. Lock Hole You can lock the instrument to a
RIGOL 1 Quick Start DS2000 User’s Guide 1-14 Front Panel Function Overview VERTICAL CH1, CH2: analog input channels. The 2 channels are marked by
1 Quick Start RIGOL DS2000 User’s Guide 1-15 Decode1, Decode2: decoding function keys. Press the corresponding key to open the decoding functio
RIGOL 1 Quick Start DS2000 User’s Guide 1-16 TRIGGER MODE: press this key to switch the trigger mode to Auto, Normal or Single and the correspondin
1 Quick Start RIGOL DS2000 User’s Guide 1-17 SINGLE Press this key to set the trigger mode to “Single”. In single trigger mode, press FORCE t
RIGOL DS2000 User’s Guide II Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the i
RIGOL 1 Quick Start DS2000 User’s Guide 1-18 Knob Adjust waveform brightness: In non-menu-operation mode (menu is hidden), turn this knob to adjust
1 Quick Start RIGOL DS2000 User’s Guide 1-19 MENU Measure: press this key to open the measurement setting menu. You can set the measurement s
RIGOL 1 Quick Start DS2000 User’s Guide 1-20 Record Record: press this key to start recording the waveform. The backlight is illuminated in red
1 Quick Start RIGOL DS2000 User’s Guide 1-21 User Interface DS2000 provides 8.0 inches, WVGA (800*480) 160,000 color TFT LCD. What is worth me
RIGOL 1 Quick Start DS2000 User’s Guide 1-22 2. Channel Label/Waveform Different channels are marked by different colors and the color of the wavefo
1 Quick Start RIGOL DS2000 User’s Guide 1-23 Note: in slope trigger, runt trigger and windows trigger, there are two trigger level labels ( and
RIGOL 1 Quick Start DS2000 User’s Guide 1-24 Denote that the current menu has a lower level menu. Press this key to return to the previous menu.
1 Quick Start RIGOL DS2000 User’s Guide 1-25 To Use the Security Lock If needed, you can use the security lock (please buy it yourself) to loc
RIGOL 1 Quick Start DS2000 User’s Guide 1-26 To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the
2 To Set the Vertical System RIGOL DS2000 User’s Guide 2-1 2 To Set the Vertical System The contents of this chapter: To Enable the Chan
RIGOL DS2000 User’s Guide III Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered. Do Not Oper
RIGOL 2 To Set the Vertical System DS2000 User’s Guide 2-2 To Enable the Channel DS2000 provides 2 analog input channels (CH1 and CH2) and provides
2 To Set the Vertical System RIGOL DS2000 User’s Guide 2-3 Channel Coupling Set the coupling mode to filter out the undesired signals. For exa
RIGOL 2 To Set the Vertical System DS2000 User’s Guide 2-4 Probe Ratio You can set the probe attenuation ratio manually. The probe ratio values avai
2 To Set the Vertical System RIGOL DS2000 User’s Guide 2-5 Waveform Invert When waveform invert is enabled, the waveform display rotates 180 d
RIGOL 2 To Set the Vertical System DS2000 User’s Guide 2-6 Vertical Expansion When using VERTICAL SCALE to change the vertical scale of the analog
2 To Set the Vertical System RIGOL DS2000 User’s Guide 2-7 Channel Label You can modify the labels used to mark the analog channels (CH1 and C
RIGOL 2 To Set the Vertical System DS2000 User’s Guide 2-8 Delay Calibration When using an oscilloscope for actual measurement, the transmission del
3 To Set the Horizontal System RIGOL DS2000 User’s Guide 3-1 3 To Set the Horizontal System The contents of this chapter: Delayed Sweep
RIGOL 3 To Set the Horizontal System DS2000 User’s Guide 3-2 Delayed Sweep Delayed sweep can be used to enlarge a length of waveform horizontally to
3 To Set the Horizontal System RIGOL DS2000 User’s Guide 3-3 In delayed sweep mode, the screen is divided into two display areas as shown in th
RIGOL DS2000 User’s Guide IV Safety Terms and Symbols Terms in this Manual. These terms may appear in this manual: WARNING Warning statements ind
RIGOL 3 To Set the Horizontal System DS2000 User’s Guide 3-4 Time Base Mode Press MENU in the horizontal control area (HORIZONTAL) at the front pane
3 To Set the Horizontal System RIGOL DS2000 User’s Guide 3-5 X-Y Mode In this mode, the oscilloscope changes the two channels from voltage-tim
RIGOL 3 To Set the Horizontal System DS2000 User’s Guide 3-6 monitor the input and output signals of the circuit. Application example: measure the
3 To Set the Horizontal System RIGOL DS2000 User’s Guide 3-7 Note: In Y-T mode, the oscilloscope could use any sample rate (within the guara
RIGOL 3 To Set the Horizontal System DS2000 User’s Guide 3-8 Roll Mode In this mode, the waveform scrolls from the right to the left to update the d
3 To Set the Horizontal System RIGOL DS2000 User’s Guide 3-9 Horizontal Scale Being similar to “Vertical Scale”, the horizontal scale can be a
RIGOL 3 To Set the Horizontal System DS2000 User’s Guide 3-10 Horizontal Reference Horizontal reference is the reference position according to which
4 To Set the Sample System RIGOL DS2000 User’s Guide 4-1 4 To Set the Sample System The contents of this chapter: Acquisition Mode Sam
RIGOL 4 To Set the Sample System DS2000 User’s Guide 4-2 Acquisition Mode The acquisition mode is used to control how to generate waveform points f
4 To Set the Sample System RIGOL DS2000 User’s Guide 4-3 number of averages as the power function of 2. The Waveform before Average: The Wav
RIGOL DS2000 User’s Guide V Measurement Category Measurement Category DS2000 series digital oscilloscopes can make measurements in Measurement C
RIGOL 4 To Set the Sample System DS2000 User’s Guide 4-4 Peak Detect In this mode, the oscilloscope acquires the maximum and minimum values of the s
4 To Set the Sample System RIGOL DS2000 User’s Guide 4-5 Sample Mode This oscilloscope only supports real-time sample. In this mode, the oscil
RIGOL 4 To Set the Sample System DS2000 User’s Guide 4-6 Sample Rate The sample rate of this oscilloscope is up to 2 GSa/s. Note that the sample rat
4 To Set the Sample System RIGOL DS2000 User’s Guide 4-7 3. Waveform Leakage: when the sample rate is too low, the waveform rebuilt from the s
RIGOL 4 To Set the Sample System DS2000 User’s Guide 4-8 Memory Depth Memory depth refers to the number of waveform points that the oscilloscope can
4 To Set the Sample System RIGOL DS2000 User’s Guide 4-9
RIGOL 4 To Set the Sample System DS2000 User’s Guide 4-10 Antialiasing At slower sweep speed, the sample rate is reduced and a dedicated display alg
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-1 5 To Trigger the Oscilloscope For trigger, you set certain trigger condition accor
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-2 Trigger Source Press MENU Source in the trigger control area (TRIGGER) at the front pa
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-3 Trigger Mode Trigger mode affects the way in which the oscilloscope searches for t
RIGOL DS2000 User’s Guide VI Ventilation Requirement This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust ar
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-4 Auto: No matter whether the trigger condition is met, there is always waveform display. A
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-5 Trigger Coupling Trigger coupling decides which kind of components will be transmi
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-6 Trigger Holdoff Trigger holdoff can be used to stably trigger the complex waveforms (suc
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-7 Noise Rejection Noise Rejection adds additional hysteresis to the trigger circuitr
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-8 Trigger Type DS2000 provides various trigger functions, including various serial bus tri
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-9 Edge Trigger Trigger on the trigger threshold of the specified edge of the input s
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-10 turning the knob, the trigger level line and the trigger mark disappear in about 2 s.
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-11 Pulse Trigger Trigger on the positive or negative pulse with a specified width.
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-12 When the Pulse Condition is set to , , or , press Setting and use to inp
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-13 Runt Trigger This trigger mode is used to trigger pulses that pass through one tr
RIGOL DS2000 User’s Guide VII Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -20℃ to +70℃ Humidity Under +35℃: ≤90% relati
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-14 <: trigger when the runt pulse width is lower than the upper limit of pulse width.
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-15 Trigger Setting: Press Setting to set the trigger parameters (trigger holdoff and
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-16 Windows Trigger (Option) Windows trigger provides a high trigger level and a low trigge
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-17 Vertical Window: Press Vertical to select the desired vertical window type. For de
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-18 Nth Edge Trigger (Option) Trigger on the nth edge that appears after the specified id
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-19 Edge Number: Press Edge to set the value of “N” in Nth edge trigger and the range
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-20 Slope Trigger Trigger on the positive or negative slope of specified time. Trigger Ty
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-21 When the Slope Condition is set to , , or , press Time and use t
RIGOL 5 To Trigger the Oscilloscope DS2000 User’s Guide 5-22 is selected. : only adjust the upper limit of the trigger level. During the adjustmen
5 To Trigger the Oscilloscope RIGOL DS2000 User’s Guide 5-23 Video Trigger (HDTV Option) Trigger on the standard video signal field or line of
Comentarios a estos manuales