RIGOL Chapter 1 Quick Start
1-16 MSO2000A/DS2000A Users Guide
CLEAR
Press this key to clear all the waveforms on the screen. If
the oscilloscope is in the RUNstate, new waveforms
will still be displayed.
RUN/STOP
Press this key to set the state of the oscilloscope to
RUNor STOP.
In the RUNstate, the key is illuminated in yellow.
In the STOPstate, the key is illuminated in red.
SINGLE
Press this key to set the trigger mode to Singleand this
key is illuminated in orange. In single trigger mode, press
FORCE to generate a trigger signal immediately.
AUTO
Press this key to enable the waveform auto setting
function. The oscilloscope will automatically adjust the
vertical scale, horizontal time base and trigger mode
according to the input signal to realize optimum
waveform display.
Note: Auto setting requires that the frequency of the
signal under test should be no lower than 20Hz if the
signal under test is Sine. Otherwise, the waveform auto
setting function may be invalid and the quick parameter
measurement function displayed in the menu will also be
unavailable.