User’s Guide RIGOL Publication number: UGC01106-1110 DM3000 Series Digital Multimeter DM3061/2/4 DM3051/2/4
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series VIII Content Safety Notices ...
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-50 Press Æ Log Æ Stop, the interface shows below. Figure 2- 59 The Da
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-52 2. The Data Log Trigger and Delay To set the start condition and the
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-54 3. The Data log stop To set the conditions to stop Data Log. Press
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-56 Multi-Route Scanning Choose , the menu of High-speed data sampling is
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-58 To set up the name of the Scanning task Press Name, and the input inter
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-60 Press Done, choose Save or Apply to set the needed task. Figure 2- 75
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-62 2.To Edit the Scanning task Press Edit, enter the task editing interf
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-64 How to Use the Built-in Help System The built-in help system provides
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-4 4. Start Statistical measurement. (1). Press ON, turn on the Statistic
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-6 Example 3: dBm Measurement Do these steps as follow: 1. Connect test le
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-8 8. Lead test leads into circuit and start to measure. Figure 3- 8
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-10 Example 6: Temperature Sensor Setting a sensor is the same way of setti
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-12 c) Press Unit Æ ℃, select the unit of the sensor: ℃. Figure 3- 20 Wh
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 3-14 b) Current sensor: Figure 3- 26 Sensor
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 4-2 12. DCV is -1200V to 1200V In limit test and null value settings the DC
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 4-4 To Change the Electric Power Fuse The electric power fuse located in the
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-1 Chapter 5 Characteristics Char
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-2 [1] Specifications are for 60 minute warm–up and select resolving index
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-3 AC Characteristics Accuracy Spe
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-4 [1] Specifications are for 60 minute warm–up and select resolving index
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-5 Frequency Period Characteristic
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-2 General Inspection Inspect a new DM3000 Digital Multimeter with the fol
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-6 Capacitance Characteristics Accuracy Specifications (% of reading + % of
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-7 Measuring Characteristics DC Vol
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-8 Shunt Resistor: 0.025Ω for 1A, 10A, 1.025Ω for 200mA, 11.025Ω for 20m
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-9 Inspection Function on Rear Pan
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-10 General Specifications Display: 256 x 64 pixels
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-11 Characteristics for DM305x DC C
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-12 [1] Specifications are for 60 minute warm–up and select resolving inde
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-13 AC Characteristics Accuracy Sp
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-14 [1] Specifications are for 60 minute warm–up and select resolving inde
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-15 Frequency Period Characteristic
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-16 Capacitance Characteristics Accuracy Specifications (% of re
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-17 Measuring Characteristics DC Vo
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-18 Shunt Resistor: 0.025Ω for 1A, 10A, 1.025Ω for 200mA, 11.025Ω for 20
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-19 Inspection Function on Rear P
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 5-20 General Specifications Display: 256 x 64 pixels
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 6-1 Chapter 6 Appendix Appendix A
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 6-2 Appendix B: Warranty RIGOL warrants that the products that it manufactu
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 6-3 Appendix C: General Care and Cle
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 6-4 Appendix D: Contact RIGOL If you have any problem or requirement occurs w
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-4 The Front/Rear Panel and User Interface It is important to get familiar
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-6 To Measure DC Voltage The following shows the system connections and se
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-8 To Measure AC Voltage The following shows the system connections and se
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-10 To Measure DC Current The following shows the system connections and s
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-12 To Measure AC Current The following shows the system connections and s
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-14 To Measure Resistance The following shows the system connections and s
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-16 4-Wire Resistance Measurement Figure 1- 21 Table1- 6 Resistance
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-18 To Measure Capacitance The following shows the system connections and
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-20 To Test Continuity The following shows the system connections and the
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-22 To Check Diodes The following shows the system connections and selecti
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-24 To Measure Frequency and Period The following shows the system connect
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-26 Period Test Figure 1- 34 Period Measurement Data Interface
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series II Safety Notices Review the following safety precautions
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-28 To Measure Sensor The DM3000 converts the sensor physical properties i
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-30 Press Add button, to input the Measured and Corresponding value to the
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-32 Figure 1- 53 The History Data Use the history function to review or s
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-34 To Choose Reading Resolution The measurement reading resolutions (the
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-36 To Choose Range Options Use “manual” selection or “automatic” to choos
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 1-38 To Control Trigger Options Use or to trigger the multimeter. When
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-2 To Set up Measurement Parameters Press key to enter the Measurement
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-4 AC Filter There are three settings at the AC Filter menu. Choose the app
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-6 Null Measurement The DM3000 null settings is available for DC voltage, A
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-8 Ratio Measurement Use Ratio measurement to measure the ratio of 2 DC vol
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series IV Safety Terms and Symbols Terms in This Guide. These ter
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-10 Frequency Measurement Frequency measurement function is used for measur
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-12 Math Functions Press key, the display shows: Figure 2- 11 The DM30
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-14 Math Functions Selective The DM3000 provides five math functions: Null
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-16 2. Limit Measurement Use the Limit test function to perform pass/fail
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-18 4. dBm Measurement This function applies to AC voltage and DC voltage
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-20 Auto Triggering Auto triggering takes continuous measurement at the hig
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-22 Reading Hold Function Start the Reading Hold Function, the hold measure
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-24 External Triggering is used to set the parameter which initiates the
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-26 To Set up the VMC At external triggering mode, when the data sampling i
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-28 Local/U-Disk Storage Local storage block is built-in the multimeter. Th
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series VI RIGOL DM3000-Series Digital Multimeter is a equipment
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-30 Figure 2- 30 NOTE Store, recall and delete use the same interface.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-32 To Set Up the Utility The Utility function establishes system paramete
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-34 Set Up LAN I/O Parameter LAN Parameters Following the procedures to se
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-36 DNS Settings Press Æ I/O Æ LAN Æ DNS, the display shows: Figure 2-
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-38 Set Up RS-232 I/O Parameters Choose RS-232 (serial data interface) inte
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-40 Select languages DM3000 supports two kinds of languages for users. Pres
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-42 Beeper On/Off Press Æ Sys Æ Sound, enter the menu shown below: Fig
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-44 Set Up Digit Format Press Æ Sys Æ Format, and enter the menu shown
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-46 Factory Defaults Setting Press Æ System Æ Cfg Æ Default, then restart
RIGOL © 2007 RIGOL Technologies, Inc.
RIGOL © 2007 RIGOL Technologies, Inc. User’s Guide for DM3000 Series 2-48 High-Speed Data Log High-speed data log features: display mode setting
RIGOL © 2007 RIGOL Technologies, Inc.