RIGOL Command Systems
Programming Guide for DS1000B Series
2-14
:ACQuire:AVERages <count>
:ACQuire:AVERages?
Function:
The command sets the average acquisition time. The <count> range is 2~256,
and the count increases by the power operation of 2.
Returned Format:
The query returns 2 or 4, 8, 16, 32, 64, 128, 256.
Example:
:ACQ:AVER 16 Set the average acquisition time as 16.
:ACQ:AVER? Return 16.
4. :ACQuire:SRATe?
Command Format:
:ACQuire:SRATe? [{<CHANnel<n>}]
Function
To query sample rate of CHANnel <n>, <n> may be 1, 2, 3, 4.
Returned Format:
The query returns 5.000e005, the unit is Sa/s.