Rigol DS1000E series Guía de usuario Pagina 99

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 162
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 98
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-71
Pass/Fail
The Pass/Fail function monitors changes of signals and output pass or fail signals by
comparing the input signal that is within the pre-defined mask. The testing results
not only can be displayed on screen or output by isolated pass/fail port but also can
be alarmed if turn on system sound.
Press UtilityPass/Fail to go to the following menu.
Figure 2- 94 Table 2- 58 The Pass/Fail Menu (Page 1/2)
Menu
Setting
Comments
Enable Test
ON
OFF
Turn on Pass/Fail test
Turn off Pass/Fail test
Source
CH1
CH2
Select Pass/Fail test on CH1
Select Pass/Fail test on CH2
Operate
(RUN)
(STOP)
Pass/Fail test stopped, press to run
Pass/Fail test running, press to stop
Msg display
ON
OFF
Turn on Pass/Fail information
display
Turn off Pass/Fail information
display
Figure 2- 95 Table 2- 59 The Pass/Fail Menu (Page 2/2)
Menu
Settings
Comments
Output
Fail
Fail +
[1]
Pass
Pass+
Output when Fail condition
detected.
Output and beep when Fail
condition detected.
Output when Pass condition
detected.
Output and beep when Pass
condition detected.
Stop
on Output
ON
OFF
Stop test when output occur
Continue test when output occur
Mask Setting
Go to mask setting menu
NOTE
[1]
: The beeper should be opened.
Vista de pagina 98