To do Track mode Cursor Measurements, follow these steps:
1. Select the Track mode for cursor measurement by pressing soft button as:
CursorModeTrack.
2. Select channel Source for Cursor A and Cursor B by pressing soft button as:
CursorCursor A or Cursor BCH1, CH2 or None.
3. Move the cursors to adjust the horizontal positions of the cursors: (Details in the
following Table)
Table 2- 87 The Cursor usage
Cursor
Operation
Cursor A
Turn the multifunctional knob ( ) to move cursor A horizontally
Cursor B
Turn the multifunctional knob ( ) to move Cursor B horizontally
NOTE: Moving cursor horizontally is not allowed when other (not tracking cursor)
menu is activated
4. To get measurement value:
Position of Cursor 1 (Time cursor centered on the midpoint of screen; Voltage
cursor centered on channel ground level).
Position of Cursor 2 (Time cursor centered on the midpoint of screen; Voltage
cursor centered on channel ground level).
Read the horizontal space between Cursor 1 and 2 (X): Time between cursors,
units in seconds.
(1/X), units in Hz, kHz, MHz, GHz.
Vertical space between cursor 1 and 2 (Y): Voltage between cursors, units in V.