Rigol M300 Manual de usuario Pagina 108

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 144
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 107
RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
3-4
Example 2: Elimination Leads Resistance Error
When measuring smaller resistance, the test leads resistance may causes large
measurement deviation.
Do these steps as follow:
1. To measure a Resistance. Connect test leads as shown in Figure 1- 19.
2. Press button, select Resistance (2WR) measurement function. Choose an
appropriate measurement range.
3. Set the Null measurement function parameters.
(1). Short circuit the test leads.
(2). Press Null Current to set the Null setting value with current
reading.
(3). Press save this setting.
(4). Press to finish the setting and back to a higher level menu.
4. In resistance measurement display interface, press Null, start Null function.
5. Lead test leads into circuit and start to measure.
Figure 3- 2 Null Measurement
Vista de pagina 107
1 2 ... 103 104 105 106 107 108 109 110 111 112 113 ... 143 144

Comentarios a estos manuales

Sin comentarios