RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DS1000/DS1000CA Series
2-46
Example:
:TIMebase:MODEMAIN Set the main timebase
:TIMebase:OFFSet1 Set the timebase offset as 1s
:TIMebase:OFFSet? Query returns 1.000e+00
3. :TIMebase[:DELayed]:SCALe
Command Format:
:TIMebase[:DELayed]:SCALe<scale_val>
Explanation:
The command is to set the time base scale in MAIN or DELAYE mode. The unit
is “S/div”. When the Delayed is ON, change the Delayed Timebase Scale will
change the width of window to amplify waveform.
<scale_val> is 2ns - 50s when it is Normal mode
500ms - 50s when it is Roll mode
When MAIN is ON, omit [: DELayed].
Query Format:
:TIMebase[:DELayed]:SCALe?
Returned Format:
Query returns the set value of < scale_val> in scientific numeric notation.
Example:
:TIMebase:MODEDELayed Set the MAIN timebase
:TIMebase:SCALe2 Set the timebase scale as 2s
:TIMebase:SCALe? Query returns 2.000e+00
4. :TIMebase:FORMat
Command Format:
:TIMebase:FORMat<value>
Explanation:
The command is to set the timebase format as XY, YT or SCANning.
Query Format: