Specifications
All specifications apply to the DS1000CA Series Oscilloscopes unless noted otherwise. To meet these
specifications, two conditions must first be met:
The instrument must have been operating continuously for thirty minutes within the specified
operating temperature.
Must perform Self Calibration operation, accessible through the Utility menu, if the operating
temperature changes by more than 5℃.
All specifications are guaranteed unless noted “typical”.
Specifications
2GSa/s (single channel)
[1]
1GSa/s (each channel)
A waveform will be displayed one time while all the channels finish N times
Sample, N could be selectable from 2, 4, 8, 16, 32, 64, 128 and 256
1MΩ± 2%, in parallel with 15pF± 3pF
50Ω± 2%
[3]
Probe Attenuation Factors
1X, 5X, 10X, 50X, 100X, 500X, 1000X
300V (DC+AC Peak, 1MΩ input impedance, 10X)
5V (DC+AC Peak, 50Ω input impedance, BNC)
[3]
Time Delay between Channel (typical)
1Sa/s-2GSa/s (Real-Time), 50GSa/s (Equivalent)
[2]
Up to 10k samples for single channel
5k samples for each channel
Scanning Speed Range
(Sec/div)
1ns/div-50s/div, DS1302CA
2ns/div-50s/div, DS1102CA, DS1202CA
5ns/div-50s/div, DS1062CA
1-2-5 Sequence
Sample Rate and
Delay Time Accuracy
± 50ppm (any time interval≥1ms)
Delta Time
Measurement Accuracy
(Full Bandwidth)
Single-shot: ± (1 sample interval + 50ppm × reading + 0.6 ns)
>16 averages: ± (1sample interval + 50ppm × reading + 0.4 ns)
Voltage difference between cursors (∆V)
Time difference between cursors (∆T)
Reciprocal of ∆T in Hertz (1/∆T)