RIGOL Command Systems
Programming Guide for DS1000E, DS1000D Series
2-60
1. :MEASure:CLEar
Command Format:
:MEASure:CLEar
Function Explanation:
The command clears the current parameter value.
2. :MEASure:VPP?
Command Format:
:MEASure:VPP? [<source>]
Function Explanation:
The command queries the peak-peak value of the waveform under measure.
<source> could be CHANnel1 or CHANnel2.
Returned Format:
Such as: 5.12e+03, the unit is V.
3. :MEASure:VMAX?
Command Format:
:MEASure:VMAX? [<source>]
Function Explanation:
The command queries the maximum of the waveform under measure.
<source> could be CHANnel1 or CHANnel2.
Returned Format:
Such as: 2.60e+03, the unit is V.
4. :MEASure:VMIN?
Command Format:
:MEASure:VMIN? [<source>]